Extended sensitivity for the calcium selective electrode
โ Scribed by Richard W. Morton; Jun Ki Chung; James L. Miller; J.Paul Charlton; Roger S. Fager
- Book ID
- 102987544
- Publisher
- Elsevier Science
- Year
- 1986
- Tongue
- English
- Weight
- 605 KB
- Volume
- 157
- Category
- Article
- ISSN
- 0003-2697
No coin nor oath required. For personal study only.
โฆ Synopsis
Sensitivity of calcium-selective electrodes heretofore has been limited to calcium concentrations above 10(-8) M in the absence of competing ions. We describe the use of calcium buffers to stabilize the free calcium in the reference electrode. Electrode calibration is linear to 10(-8) M and is curvilinear to 10(-11) M in the presence of 0.1 M ionic strength. Selectivity with respect to competing cations, magnesium, potassium, sodium, and hydrogen is preserved. Electrode response time is less than 2 s for small changes in calcium activity. Response range is linear over 9 log units of calcium activity. Potential-time stability is less than 10 mV/h at saturation currents. Although the silver-silver chloride terminals are photosensitive throughout the visible and near-ultraviolet regions, housing the reference and indifferent in opaque barrels avoids false photovoltaic response.
๐ SIMILAR VOLUMES