Impedance studies on test circuits demonstrated that three-point measurements are susceptible to voltage divider distortion/artifacts if the reference electrode impedance approaches the analyzer input impedance. These features can be avoided by using an instrument with a high input impedance or a re
✦ LIBER ✦
Experimental limitations in impedance spectroscopy: Part II — electrode artifacts in three-point measurements on Pt/YSZ
✍ Scribed by G. Hsieh; T.O. Mason; L.R. Pederson
- Book ID
- 104296510
- Publisher
- Elsevier Science
- Year
- 1996
- Tongue
- English
- Weight
- 828 KB
- Volume
- 91
- Category
- Article
- ISSN
- 0167-2738
No coin nor oath required. For personal study only.
✦ Synopsis
High temperature (570-750°C) three-point impedance measurements on the system Pt/YSZ using a platinum reference electrode exhibited distortions/artifacts due to the voltage divider effect. These distortions could be avoided by using a lower impedance reference electrode (Ag) or employing an instrument with a higher input impedance. A correction procedure was also successful.