Experimental evaluation of ALFARD treatment planning system for 6 MV photon irradiation: a lung case study

✍ Scribed by Asghar Mesbahi; Mahmoud Allahverdi; Hossein Gheraati; Ehsan Mohammadi


Publisher
Wielkopolskie Centrum Onkologii
Year
2004
Tongue
Polish
Weight
322 KB
Volume
9
Category
Article
ISSN
1507-1367

No coin nor oath required. For personal study only.

✦ Synopsis


Dooewiadczalna ocena systemu planowania leczenia na przyk³adzie naoewietlania p³uc fotonami o energii 6 MV

Streszczenie

Purpose: Simple inhomogeneity correction methods available in a number of currently applied treatment planning systems are not accurate enough for dose calculations in lung irradiations. The purpose of this study was to evaluate the accuracy of the ALFARD treatment planning system in dose calculations for lung irradiation. Material and methods: An anatomic thorax phantom and a 6 MV photon beam were used for our irradiation. Our set-up consisted of an an-22 terior single field for the left lung of a thorax phantom with field sizes of 5 x 5 cm and 10 x 10 cm . The percentage depth doses for each point in the lung were measured by a Pinpoint ionization chamber and calculated by the ALFARD treatment planning system. The results of calculations and measurements were compared.

Results:

The ALFARD calculations overestimated measurements at all points and field sizes. The magnitude of error increased with depth 22 of the calculation point from 2.7% to 17.3% for the field size of 5 x 5 cm . The error for 5 x 5 cm was approximately twice as high as that 2 for 10 x 10 cm . Conclusions: The ALFARD treatment planning system cannot calculate the dose in the lung accurately. This may be due to inherent deficiencies of the effective path length method, which is implemented in the ALFARD treatment planning system.