Ex vivo hemodynamic assessment of mechanical aortic valve gradients using a high-fidelity pressure wire

✍ Scribed by Andrew D. Michaels; David Mester


Publisher
John Wiley and Sons
Year
2011
Tongue
English
Weight
168 KB
Volume
77
Category
Article
ISSN
1522-1946

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

Objectives: Accurate assessment of prosthetic mechanical valve malfunction is challenging for non‐invasive and invasive techniques. We evaluated a 0.014‐inch pressure‐sensing coronary guidewire to assess mechanical valve dysfunction. Background: Several case reports have shown that transaortic pressure gradients can be recorded using a 0.014‐inch guidewire during cardiac catheterization. Methods: We performed an ex vivo study measuring the effects of sequentially crossing the center of each valve with a 6 French coronary angiographic catheter, a 0.035‐inch guidewire, and a 0.014‐inch pressure‐sensing guidewire on valve dysfunction using the following 23 mm bileaflet and tilting‐disc aortic valves: St. Jude Regent™, CarboMedics, Medtronic Hall™, and Björk‐Shiley Monostut. A left heart model pulse duplicator recorded 10 consecutive cycles. Results: For all valves, the greatest increase in valve regurgitation occurred with the 6 French catheter, causing a reduction in aortic valve back pressure and cardiac output, with an increase in leakage rate, regurgitant fraction, and energy loss. In comparison to the 0.035‐inch guidewire, the 0.014‐inch guidewire had greater regurgitation for St. Jude, lower for Medtronic Hall, and equivalent for CarboMedics and Björk‐Shiley valves. For the CarboMedics valve, the 0.035‐inch guidewire caused a significant increase in regurgitant fraction and energy loss, while the pressure wire had no change compared to baseline. Conclusions: The degree of regurgitation caused by the 0.014‐inch guidewire varies with the type of mechanical aortic valve. While prior case reports have shown that valve hemodynamics may be measured using a pressure‐sensing guidewire, valve regurgitation occurs when crossing a St. Jude, Medtronic Hall, or Björk‐Shiley aortic valve. © 2011 Wiley‐Liss, Inc.


📜 SIMILAR VOLUMES