𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Evaluation of the charge-sharing effects on spot intensity in XRD setup using photon-counting pixel detectors

✍ Scribed by H.-E. Nilsson; C.G. Mattsson; B. Norlin; C. Fröjdh; K. Bethke; R. De Vries


Publisher
Elsevier Science
Year
2006
Tongue
English
Weight
336 KB
Volume
563
Category
Article
ISSN
0168-9002

No coin nor oath required. For personal study only.

✦ Synopsis


In this study, we examine how charge loss due to charge sharing in photon-counting pixels detectors affects the recording of spot intensity in an X-ray diffraction (XRD) setup. In the photon-counting configuration, the charge from photons that are absorbed at the boarder of a pixel will be shared between two pixels. If the threshold is high enough, these photons will not be counted whereas if it is low enough, they will be counted twice. In an XRD setup, the intensity and position of various spots should be recorded. Thus, the intensity measure will be affected by the setting of the threshold. In this study, we used a system level Monte Carlo simulator to evaluate the variations in the intensity signals for different threshold settings and spot sizes. The simulated setup included an 8keV mono-chromatic source (providing a Gaussian shaped spot) and the MEDIPIX2 photon-counting pixel detector (55 mm  55 mm pixel size with 300mm silicon) at various detector biases. Our study shows that the charge-sharing distortion can be compensated by numerical post processing and that high resolution in both charge distribution and position can be achieved.