๐”– Bobbio Scriptorium
โœฆ   LIBER   โœฆ

Evaluation of the Beckman Synchron cx4 clinical chemistry analyzer in a hospital laboratory

โœ Scribed by T. Ambus; N. Korogyi; F. Decampos; B. Groom; V. T. Innanen


Publisher
John Wiley and Sons
Year
1990
Tongue
English
Weight
511 KB
Volume
4
Category
Article
ISSN
0887-8013

No coin nor oath required. For personal study only.

โœฆ Synopsis


The Beckman Synchron CX4 randomaccess multianalyzer was evaluated in a medium sized hospital laboratory. The instrument does end-point, rate, and multipoint assays and carries on-board reagents for 24 tests. In addition to predefined tests, the instrument can be programmed for 100 userdefined tests; these are stored on the hard disk and can utilize up to three component reagents each. The throughput is 200 tests per hour. There is stat testing capability. In our evaluation, within-run and between-run precision and linearity were good, and no reagent carryover was detected. There was good correlation with the in-house methodology for the 19 tests evaluated. A disadvantage at the time of evaluation was interference by elevated bilirubin on creatinine, phosphorus, uric acid, and triglycerides. This problem of interference is being addressed by the manufacturer.