Evaluation of an Ultrasonic Nebulizer Interface for Capillary Electrophoresis and Inductively Coupled Plasma Mass Spectrometry
โ Scribed by Qinghong Lu; Ramon M. Barnes
- Publisher
- Elsevier Science
- Year
- 1996
- Tongue
- English
- Weight
- 161 KB
- Volume
- 54
- Category
- Article
- ISSN
- 0026-265X
No coin nor oath required. For personal study only.
โฆ Synopsis
An interface for capillary electrophoresis (CE) and inductively coupled plasma mass spectrometry (ICP-MS) is developed with a commercial ultrasonic nebulizer. The CE ground path was provided through a coaxial liquid electrolyte flow surrounding the CE capillary. To accommodate this dual capillary arrangement, the ultrasonic nebulizer spray chamber sample uptake tube inner diameter was enlarged. Metallothioneins were separated and detected to characterize the interface and to compare it with a concentric-glass nebulizer interface. Capillary electrophoresis resolution was unaffected by the ultrasonic nebulizer interface. Signal intensity was improved 3.6-to 8.4fold compared to the concentric nebulizer arrangement, while signal-to-noise ratio improvement was only 1.3-to 2.8-fold. Concentration detection limits for 114 Cd, 111 Cd, 64 Zn, and 66 Zn are 0.082, 0.039, 1.18, and 1.15 ppm in metallothionein, while the corresponding absolute detection limits are 2.5, 1.2, 37, and 36 fg, respectively.
๐ SIMILAR VOLUMES
## Abstract A capillary electrophoresisโinductively coupled plasmaโmass spectrometric (CEโICPโMS) method for the speciation of six arsenic compounds, namely arsenite [As(III)], arsenate [As(V)], monomethylarsonic acid, dimethylarsinic acid, arsenobetaine and arsenocholine is described. The separati