## Abstract The k‐space trajectory of a spiral imaging sequence was measured with a self‐encoding technique. The image quality improved dramatically when reconstructed with the measured k‐space trajectory. There were substantial artifacts in images reconstructed with the derived k‐space trajectory
Estimation of k-space trajectories in spiral MRI
✍ Scribed by Hao Tan; Craig H. Meyer
- Publisher
- John Wiley and Sons
- Year
- 2009
- Tongue
- English
- Weight
- 643 KB
- Volume
- 61
- Category
- Article
- ISSN
- 0740-3194
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
For non‐Cartesian data acquisition in MRI, k‐space trajectory infidelity due to eddy current effects and other hardware imperfections will blur and distort the reconstructed images. Even with the shielded gradients and eddy current compensation techniques of current scanners, the deviation between the actual k‐space trajectory and the requested trajectory remains a major reason for image artifacts in non‐Cartesian MRI. It is often not practical to measure the k‐space trajectory for each imaging slice. It has been reported that better image quality is achieved in radial scanning by correcting anisotropic delays on different physical gradient axes. In this article the delay model is applied in spiral k‐space trajectory estimation to reduce image artifacts. Then a novel estimation method combining the anisotropic delay model and a simple convolution eddy current model further reduces the artifact level in spiral image reconstruction. The root mean square error and peak error in both phantom and in vivo images reconstructed using the estimated trajectories are reduced substantially compared to the results achieved by only tuning delays. After a one‐time calibration, it is thus possible to get an accurate estimate of the spiral trajectory and a high‐quality image reconstruction for an arbitrary scan plane. Magn Reson Med, 2009. © 2009 Wiley‐Liss, Inc.
📜 SIMILAR VOLUMES
## Abstract Spiral acquisitions are used in fast cardiac imaging because they traverse __k__‐space efficiently and minimize flow artifacts. A variable pitch logarithmic spiral trajectory is designed to critically sample the low‐frequency region in __k__‐space and gradually undersample the high‐freq
## Abstract Dynamic cardiac imaging in MRI is a very challenging task. To obtain high spatial resolution, temporal resolution, and signalto‐noise ratio (SNR), single‐shot imaging is not sufficient Use of multishot techniques resolves this problem but can cause motion artifacts because of data incon
## Abstract Timing delays between data acquisition and gradient transmission result in image degradation. This is especially true in spiral MRI, where delays can alter data in a nonuniform manner, generating significant artifact in the reconstructed data. The many methods that exist to mitigate the
A method is presented to correct for the imperfections of spatial encoding gradients in MRI. The approach is simple and fast, can be performed with standard scanner hardware, and does not require separate measurements with reference phantoms. The new method, using the MR signal to accurately measure