Errors in T1-determination using multislice technique and Gaussian slice profiles
โ Scribed by M. Just; H.P. Higer; P. Pfannenstiel
- Book ID
- 103909155
- Publisher
- Elsevier Science
- Year
- 1988
- Tongue
- English
- Weight
- 890 KB
- Volume
- 6
- Category
- Article
- ISSN
- 0730-725X
No coin nor oath required. For personal study only.
โฆ Synopsis
Multislice SE-or IR-sequences with different TRand T,-times, respectively, are often used, to determine Tland T,-values for tissue characterisation. To investigate the perturbating influence of the sequential excitation in multislice technique, we measured the signal intensities and calculated the T,-values of phantoms as a function of gap widths between neighboured slices. Phantoms contained fluids of different T,. We found a strong dependence of signal intensities and therefore of calculated T,-values on the gap width, when a T,/T,-ratio of more than 0.3-0.4 was reached. T,-values are considerably overestimated in this case, whereas T,-values are not influenced by the sequential excitation. We conclude that reliable T,-measurements necessitate a large spacing of more than 1 slice thickness between adjacent slices.