The fatigue resistance of a single-lap aluminium adhesive joint to cyclic loading in combined shear and bending mode is investigated by nonlinear finite element analysis and crack propagation experiments. The epoxy adhesive is modelled by an elasto-plastic overlay material model. The initial cycles
Environmental fatigue testing of silastic finger joint prostheses
β Scribed by Weightman, B. ;Simon, S. ;Rose, R. ;Paul, I. ;Radin, E.
- Publisher
- John Wiley and Sons
- Year
- 1972
- Tongue
- English
- Weight
- 645 KB
- Volume
- 6
- Category
- Article
- ISSN
- 0021-9304
No coin nor oath required. For personal study only.
β¦ Synopsis
Heat vulcanized silicone elastomer (Silastic-Dow) has been in clinical use as a prosthetic replacement material for over a decade. When subjected to operating stress in body fluids, as in cardiac poppet-valves, failure has occurred. Fractures of silastic finger-joint prosthesis have also been experienced.
Chemical and fractographic analyses of such failed finger-joint prostheses raise the suspicion of fatigue failure. Severe discoloration of the implant at the point of maximal stress concentration, the presence of significant amounts of triglyceride and cholesterol on the fractured prostheses, and the pattern of fracture markings indicate that fracture was due to a combination of chemical surface deterioration and stress concentration, Study of the method of surgical insertion of these prostheses suggests failure to completely excise the volar plate and other palmar joint structures could create a s t e p off and sufficiently increase the stress in the bending element of the device to cause fracture. In vitro cyclic testing a t 37Β°C in serum failed to cause fracture of these devices after 10 million cycles, but did produce gradual discoloration of the entire prosthesis which began and was most noticeable at the point of bending. Although the prostheses adsorb lipid, which may produce surface deterioration, it is anticipated that properly inserted, the service life of these devices should be more than adequate for full clinical application.
π SIMILAR VOLUMES
The fatigue strength of transverse non-load-carrying cruciform welded joints of SMSOB steel has been investigated by R = 0 test and omrr = oy test which simulates the tensile residual stress exists in real welded structures. The plate thickness was 20 mm and 40 mm. The residual stress, or, at the we
The thickness effect on fatigue strength of transverse butt welded joints was investigated by a test method which simulates the behavior of large sized welded structures having tensile residual stresses by means of ordinary narrow specimens containing a low level of residual stresses. In the test, s