Emission characteristics of plastic syringes sterilized with ethylene oxide—A controlled study
✍ Scribed by Yeh-Chung Chien; Po-Chi Su; Lien-Hsiung Lee; Chang-Yuh Chen
- Publisher
- John Wiley and Sons
- Year
- 2009
- Tongue
- English
- Weight
- 118 KB
- Volume
- 91B
- Category
- Article
- ISSN
- 1552-4973
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Objectives:
This study examined the emission characteristics of ethylene oxide (EO)‐sterilized syringes under various environmental conditions, aiming to develop control strategies to minimize worker exposure.
Methods:
Experiments were performed in a facility in which temperature, relative humidity (RH), and air change rate (ACR) were controlled.
Results:
Analytical results indicate that the main effects of the four test variables on kinetic parameters were statistically significant (p < 0.05), except for the effect of the product on the decay rate constant, the effect of ACR on maximum EO concentration, and effect of RH on the area under the curve‐days 1 and 2. The interactive effects among test variables were also evident, indicating complex emission behaviors. The mean EO emission factors during the days 1 and 2 and at the 48th hour for the 1‐ and 30‐ml products were 2302, 1301, and 1031 mg/m^3^/h, and 871, 490, and 381 mg/m^3^/h, respectively. The times required for air EO concentrations from tested products to return to ∼0 and 1 ppm (permissible limit) were 417 and 218 h, respectively.
Conclusions:
Plastic content, temperature, RH, and ACR affected EO emissions. ACR is an achievable means of control; however, the aeration area/system should be isolated to ensure adequate ventilation is achieved. © 2009 Wiley Periodicals, Inc. J Biomed Mater Res Part B: Appl Biomater, 2009