Gas chromatography: Elution temperature,
โ
W.G. Jennings; S. Adam
๐
Article
๐
1975
๐
Elsevier Science
๐
English
โ 466 KB
The elution temperatures and degree of separation of a model system series of hydrocarbons were followed as functions of the rate of temperature programming and carrier gas velocity. Graphs are presented that permit individual assessment of these variables. Lower elution temperatures can be achieved