𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Elimination of coupling between cylindrical transmit coils and surface-receive coils for in vivo NMR

✍ Scribed by M. Robin Bendall; Alan Connelly; Jamie M. McKendry


Publisher
John Wiley and Sons
Year
1986
Tongue
English
Weight
538 KB
Volume
3
Category
Article
ISSN
0740-3194

No coin nor oath required. For personal study only.

✦ Synopsis


Coupling between large transmit and surface-receive coils is eliminated using a lambda/4 cable, terminated with crossed diodes to ground, inserted between the receive coil and ground, and by using a cable of length a little less than lambda/2 inserted between the transmit match capacitance and the normal crossed diodes in the transmit input line.


πŸ“œ SIMILAR VOLUMES


Calibrated uncoupling of tightly coupled
✍ M. Robin Bendall πŸ“‚ Article πŸ“… 1988 πŸ› John Wiley and Sons 🌐 English βš– 1007 KB

Three known active methods of detuning rf coils have been tested by determining the efficiency of uncoupling two concentric transmit surface coils in a double-coil probe. In a comparison at 80 MHz, the method based on the use of lambda/4 cables was found to have advantages. These methods utilize the

A comparison of an inductively coupled i
✍ E. D. Wirth III; T. H. Mareci; B. L. Beck; J. R. Fitzsimmons; P. J. Reier πŸ“‚ Article πŸ“… 1993 πŸ› John Wiley and Sons 🌐 English βš– 836 KB

## Abstract A study was performed to determine whether an implanted, inductively coupled nuclear magnetic resonance (NMR) imaging spine coil could provide a significant gain in signal‐to‐noise ratio (SNR) on images of the spinal cord relative to the SNR of optimized surface coils. Implanted coils w