𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Electrogenerated chemiluminescence detection for capillary electrophoresis

✍ Scribed by S. Douglass Gilman; Charlotte E. Silverman; Andrew G. Ewing


Publisher
John Wiley and Sons
Year
1994
Tongue
English
Weight
823 KB
Volume
6
Category
Article
ISSN
1040-7685

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

Electrogenerated chemiluminescence (ECL) detection based on the reaction of luminol and H~2~O~2~ has been developed for capillary electrophoresis. The ECL response at carbon fiber and platinum wire microelectrodes has been investigated. Carbon fiber electrodes are observed to be less sensitive than platinum electrodes but provide a more stable response. The ECL response at a carbon fiber electrode is linear over three orders of magnitude for luminol with a linear correlation coefficient of 0.998. Mass detection limits of 92 amol and 260 amol have been obtained for luminol using platinum wire microelectrodes and carbon fiber microelectrodes, respectively. The influence of hydrogen peroxide concentration (in the detection buffer resrvoir) on the luminescence response has been examined and found to be dependent on electrode type. The dependence of the ECL response on the applied voltage at the microelectrode has been studied for both electrode types. Capillary electrophoresis with ECL detection has been used for analysis of amines derivatized with N‐(4‐aminobutyl)‐N‐ethylisoluminol coupled to N, N‐disuccinimidylcarbonate (ABEI‐DSC). Detection limits of 2.0 fmol and 0.96 fmol are obtained for n‐octylamine and n‐propylamine, respectively. In addition, ABEI‐DSC has been used to successfully label the tripeptide Val‐Tyr‐Val, and micellar electrokinetic capillary chromatography has been used with ECL detection for the separation of ABEI‐DSC labeled amines.


📜 SIMILAR VOLUMES


A novel method developed for acetylcholi
✍ Wei Wei; Min Wei; Xuejun Kang; Huihua Deng; Zuhong Lu 📂 Article 📅 2009 🏛 John Wiley and Sons 🌐 English ⚖ 145 KB

## Abstract A novel method for highly sensitive detection of acetylcholine in royal jelly was proposed by using CE coupled with electrogenerated chemiluminescence (ECL). Acetylcholine, which could not react with Tris(2,2′‐bipyridine)ruthenium(II) to strengthen its ECL signals, decomposed into trime

Chemiluminescence detection in capillary
✍ Willy R. G. Baeyens; Betty Lin Ling; Kazuhiro Imai; Anthony C. Calokerinos; Step 📂 Article 📅 1994 🏛 John Wiley and Sons 🌐 English ⚖ 969 KB

## Abstract Capillary electrophoresis techniques offer high plate numbers and are highly suited for the efficient separation of a wide variety of chemical components in diverse matrices. Because of the small capillary and detection cell dimensions, together with the minute volumes of samples to be

Peroxyoxalate chemiluminescence detectio
✍ Loranelle L. Shultz; Scott Shippy; Timothy A. Nieman; Jonathan V. Sweedler 📂 Article 📅 1998 🏛 John Wiley and Sons 🌐 English ⚖ 160 KB 👁 2 views

Capillary electrophoresis CE is often limited by the need to measure nanoliter-volume, low-concentration analytes with selectivity and specificity. A Ž . post-column chemiluminescence CL method is described in which the capillary effluent is collected on a membrane which has been coated with a perox