A Prasslan Blue-modified glassy carbon electrode prepared by simple adsorption exhibited excellent electrocatalytic adivlty m the oxldatlon of hydrazme m ac&c mecha A film of the perfluorosulphomc acid polymer Nafion coated on top of the Prussian Blue-mochfied glassy carbon electrode can Improve the
Electrocatalytic oxidation and flow amperometric detection of hydrazine at an electropolymerized 4-vinylpyridine/palladium film electrode
โ Scribed by Tao Li; Erkang Wang
- Publisher
- John Wiley and Sons
- Year
- 1997
- Tongue
- English
- Weight
- 386 KB
- Volume
- 9
- Category
- Article
- ISSN
- 1040-0397
No coin nor oath required. For personal study only.
โฆ Synopsis
A poly(4-viny1)pyridine (PVP)/Pd film electrode was constructed for the electrocatalytic detection of hydrazine. The preparation of the PVP/GC electrode was performed by electropolymerization of the monomer 4-vinylpyridine onto the surface of a glassy carbon electrode. Subsequently, palladium is electrodeposited onto the polymer modified electrode surface. The ion-exchange function of PVP polymer is helpful to this process in view of the tetrachlorapalladate anion. Compared with the PdGC electrode, the modified electrode displays a better mechanical stability in a flowing stream. The PVPPd film electrode exhibits higher sensitivity when detecting hydrazine with a detection limit of 0.026 ng (S/N=3).