The purge and trap (P&T) method of analysis has been interfaced with fused silica capillary column gas chromatography. This interfacing has been accomplished without splitting the P&T trap desorption carrier gas.Thus, 100% of the purged compoundsare transferred to the column. The analytes are cryofo
Effects of trapping temperature and film thickness in purge and trap with whole column cryotrapping on fused silica columns
โ Scribed by Pankow, J. F.
- Publisher
- John Wiley and Sons
- Year
- 1986
- Tongue
- English
- Weight
- 969 KB
- Volume
- 9
- Category
- Article
- ISSN
- 0935-6304
No coin nor oath required. For personal study only.
โฆ Synopsis
The effects of trapping temperature and column film thickness were investigated with respect to their ability to promote effective cryofocusing on fused silica capillary columns. The study was a further development of the purge and trap with whole column cryotrapping (P&T/WCC) method. The rates at which compounds could be thermally desorbed from a P&T unit and transferred to a column (with zero split) were first examined. A near quantitative transfer of the desorbable analytes was obtained with a 4 min, 18OoC, 20 mL/min desorption. The compounds tested included naphthalene. Columns with film thicknesses from 0.1 2 to 3.0 pm were then investigated in P&T/ WCC analyses with WCC temperatures ranging from -80 to OOC. Thetrapping tookplacefrom a transfer linegas stream initially at 175OC. The volatilities of the compounds examined varied from that of 1,l -dichloroetheneto 111,2,2-tetrachloroethane. A higher film thickness was found to ease the WCC temperature requirements. Within each column type, the warmest WCC temperatures which allowed good cryotrapping with no significant increases in peak width were: 0.25 pm, -7OOC; 1.0 pm, -5OOC; 3.0 pm, -2OOC. In addition to being quantitative, the trapping provided good chromatography.
๐ SIMILAR VOLUMES