Abstract
Four prototype cementless hip stems were tested following the ISO 7206 protocol for the assessment of the endurance properties and compared with a cast Crο£ΏCoο£ΏMo (ASTM F75) commercially available stem which was used as reference design. All the tested stems were similar in shape and size, but with some substantial differences. The first was made of forged Ti6Al4V alloy (ASTM F136). The second, made of the same material, featured a central hole intended to reduce the bending stiffness of the stem itself. The third was identical to the second but for a small tooling notch in one of the fillets of the hole. The fourth was similar to the first but had a coating of sintered titanium beads in the proximal part. All of these modifications were made to evaluate the effect of fatigue strength of intentional or unintentional features commonly found in commercial stems. The forged Ti6Al4V alloy was found to be substantially stronger than the cast ASTM F75 Crο£ΏCoο£ΏMo alloy. However, tooling notches or sintered coatings were found to dramatically reduce this strength. Thus, the Ti6Al4V alloy calls for an accurate design process, especially when complex shapes or sintered structures are required. Β© 1995 John Wiley & Sons, Inc.
π SIMILAR VOLUMES
## Abstract Bone loss in the proximal femur at the time of revision hip arthroplasty for a failed primary cemented femoral component can substantially reduce the stability of the revision stem, Use of an extendedβlength femoral component has been suggested to aid in achieving longβterm fixation; ho
Recent studies have established that a mechanism of initiation of failure of fixation of cemented femoral components is debonding of the cement-metal interface. Other studies have shown that the torsional forces induced by stair climbing and rising from a chair are very high. Thus, the interface bet
In cemented total hip prostheses, the role of creep of the occurred with all three bonding conditions, allowing addiacrylic cement (polymethyl methacrylate, [PMMA]) in in-tional subsidence of the stem and a decrease in the stress creasing or decreasing the chance of failure of the cement components
Porous-coated Ti-6A1-4V has a fatigue strength approximately one-third that of the uncoated alloy. The interfacial geometry between the porous coating and the implant substrate is notchlike, leading to stress concentrations that have been shown to be the main cause for the reduction in fatigue stren