Effect of femoral head diameter on tissue concentration of wear debris
β Scribed by Hirakawa, Kazuo ;Bauer, Thomas W. ;Hashimoto, Yasushi ;Stulberg, Bernard N. ;Wilde, Alan H. ;Secic, Michelle
- Publisher
- John Wiley and Sons
- Year
- 1997
- Tongue
- English
- Weight
- 72 KB
- Volume
- 36
- Category
- Article
- ISSN
- 0021-9304
No coin nor oath required. For personal study only.
β¦ Synopsis
Although several studies have reported the physical properties of debris particles in tissues adjacent to failed total joint implants, few have correlated the results of particle analysis with other clinical and implant variables believed to influence implant wear. We retrospectively analyzed 41 fibrous membranes (from 35 patients) adjacent to failed acetabular cups from a single manufacturer and studied the relationship between three different femoral head sizes (26, 28, and 32 mm) and the characteristics of wear debris in the adjacent tissues. All total hip prostheses consisted of modular cobalt-chromium alloy femoral heads articulating with titanium-alloy-backed ultrahigh molecular weight polyethylene (UHMWPE) acetabular components from a single manufacturer. Large femoral head diameter (32 mm) was found to correlate significantly with large particle size (diameter and surface area, p < 0.05), high tissue concentration of particles (particle volume/gram of tissue, p < 0.01), and high rate of particle production (particles volume/month, p < 0.05). The results of these quantitative assays support the findings of radiographically based clinical studies that show higher volumetric wear associated with 32 mm femoral head components.
π SIMILAR VOLUMES
## Abstract Hip resurfacing arthroplasty has become a popular alternative to conventional hip surgery. Surface replacements with bearing sizes of 55 mm (__n__ = 5) and 39 mm (__n__ = 5) were tested in a hip simulator for 15 million cycles (Mc). Wear debris was isolated from the serum lubricant and
Wear debris generated by joint implant components has been reported to activate inflammatory and immune cells. Particulate debris derived from prosthetic material induces monocytes/macrophages, lymphocytes, synoviocytes, and fibroblasts to secrete cellular products, such as cytokines, which mediate
## Abstract The impact of femoral head surface roughness on wear of Ξ³βirradiation sterilized (3 MRad in nitrogen, crosslinked) and nonsterilized (not crosslinked) UHMWPE acetabular cups has been evaluated. Gravimetric wear testing was performed in a hip joint simulator for 2 Γ 10^6^ cycles. CoCrMo