The EBI (BARD-XT, C.R. Bard, Murray Hill, NJ) stent is a new radiopaque balloon expandable coronary stent with high resistance to external radial forces. It does not shorten significantly with expansion and allows stent implantation in bifurcation lesions. A total of 28 EBI stents were implanted in
Early experience with the Bard XT stent
β Scribed by Ravat, Hasmukh ;Cannon, Adam
- Publisher
- John Wiley and Sons
- Year
- 1998
- Tongue
- English
- Weight
- 349 KB
- Volume
- 45
- Category
- Article
- ISSN
- 0098-6569
No coin nor oath required. For personal study only.
β¦ Synopsis
The Bard XT stent is a new generation balloon expandable intracoronary stent. It has several unique design advantages. Between October 1996 and November 1997, 127 Bard XT stents of various length were deployed in 93 patients with 109 lesions. According to the American College of Cardiology (ACC) and American Heart Association (AHA) classifications 7 lesions were type A, 38 were type B1, 43 were type B2 and 21 were type C [Ellis et al.: Circulation 82:1193-1202, 1990]. Stent delivery was successful in 98% of attempts. Angiographic success was achieved in 98% of 109 lesions. Procedural success was achieved in 94% of 93 patients. Minimal luminal diameter (MLD) increased from 0.91 Ψ 0.34 mm to 3.03 Ψ 0.44 mm and percentage diameter stenosis reduced from 69.1 Ψ 11.07 to 9.96 Ψ 6.81. Complications occurred in four patients. One patient had intracranial hemorrhage, one patient had subacute thrombosis and two patients died postprocedure. Patients were followed for a period of 1 to 14 months (average 7 Ψ 4 months) for major cardiac events and clinical restenosis. The Bard XT stent is a user-friendly device which provided excellent angiographic results and short-term clinical outcome in selected cases. Further study is required to evaluate effects on restenosis.
π SIMILAR VOLUMES
The Multi-Link coronary stent (Advanced Cardiovascular Systems, Santa Clara, CA) is a balloon expandable stent carved from a stainless steel cylinder and is composed of 12 corrugated rings connected by multiple links. This design gives the stent great flexibility and conformity as wail as radial str
Coating stents with autologous venous grafts has been suggested to prevent problems associated with conventional stenting, but the need for surgical vessel harvest hampered broad application. A novel synthetic coronary stent graft (CSG) overcomes this limitation by a synthetic membrane, fixed betwee
We prospectively studied 223 patients (288 lesions) who underwent elective or bail out implantation of 309 NIR stents (Scimed, Boston Scientific Corporation, Galway, Ireland). Most lesions (68.4%) had unfavorable characteristics (type B2 or C). Primary success in stent deployment was achieved in 305