In current magnetic resonance imaging (MRI), valuable information must often be discarded because the NMR signal has greater dynamic range than the analog-to-digital converter (ADC) hardware. Typically, a small set of high-intensity data points near the center of the spin echo is responsible for mos
Dynamic range compression in surface coil MRI
✍ Scribed by T Sharples; R Lufkin; B Flannigan; W Hanafee
- Book ID
- 103908838
- Publisher
- Elsevier Science
- Year
- 1986
- Tongue
- English
- Weight
- 416 KB
- Volume
- 4
- Category
- Article
- ISSN
- 0730-725X
No coin nor oath required. For personal study only.
✦ Synopsis
The largs dynamic range in signal intensity present in MR surface coil images makes proper windowing and photography difficult. By removing the low spatial frequency information due to variations in surface coil field intensity from the high spatial frequency information containing the image data a considerable compression of this Qnamic range of signal intensity is possible.
A technique to accomplish this was implemented on a digital computer for use with RR surface coil imegs data. The compression W(IS done as a post processing option after the patient seen had been completed and therefore did not alter actual scan times. Although the image signaf to noise was not altered, the image quality and ease of photography for most images was improved. Thus, digital dynamic range compression is a practical technique to aid in surface coil MRI studies.
📜 SIMILAR VOLUMES
## Abstract Recent theoretical advances in the field of compressive sampling—also referred to as compressed sensing (CS)—hold considerable promise for practical applications in MRI, but the fundamental condition of sparsity required in the CS framework is usually not fulfilled in MR images. However
## Abstract MR‐guided intravascular interventions require image update rates of up to 10 images per second, which can be achieved using parallel imaging. However, parallel imaging requires many coil elements, which increases reconstruction times and thus compromises real‐time image reconstruction.
The construction and application of eight different MRI surface coils is described. The coils consist of an anatomically shaped copper wire loop as an antenna and a printed circuit board containing electronic components for tuning and matching. The electronic device for tuning and matching is interc
## Abstract The present study describes the advantageous application of a surface coil to electron paramagnetic resonance (EPR) irradiation in Overhauser‐enhanced MRI (OMRI). OMRI is a double‐resonance method for imaging free radicals based on the Overhauser effect. Proton NMR images are recorded w