## Abstract Bicaval dual lumen catheters improve the efficiency of venoβvenous extracorporeal membrane oxygenation by minimizing recirculation with an innovative design, which requires precise placement of three catheter ports in the superior vena cava, right atrium, and inferior vena cava, respect
Dual-lumen catheters: Quality control tests for radiopacity
β Scribed by Kimberlee J. Kearfott
- Publisher
- Elsevier Science
- Year
- 1991
- Weight
- 928 KB
- Volume
- 42
- Category
- Article
- ISSN
- 0883-2889
No coin nor oath required. For personal study only.
β¦ Synopsis
A protocol utilizing a single-point densitometer was developed for the quality control (Q.C.) of the radiopacity of dual-lumen radiographic catheters where x-rays generated at 80 kVp are typically used. Catheter contrast, defined as the difference between catheter radiopacity and background optical density (O.D.) when an aluminum plate is imaged, forms the basic Q.C. parameter. When a given catheter is repeatedly measured, catheter contrast varies with a standard deviation of 0.013 O.D. units. The effects of minor background O.D. variations on contrast are significant, but empirically correctable. A variation in measured contrast of approx. 0.01 O.D. units is observed per 10 kVp energy variation. Aluminum plate thickness has a significant effect on the results; however, other details of the protocol geometry are less important. Variations due to catheter orientation were examined and imaging of packaged catheters was demonstrated.
π SIMILAR VOLUMES
The Western Turkey Mesonet was initiated in 2002 as part of the Turkey Emergency Flood and Earthquake Recovery project. Its main goal is to provide agricultural and meteorological data not only to support flood forecasting/warning applications but also meteorological uses such as nowcasting and/or s