𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Discrimination of ventricular tachycardia from sinus tachycardia by antitachycardia devices: value of median filtering

✍ Scribed by Chih-ming James Chiang; Janice M. Jenkins; Lorenzo A. DiCarlo


Publisher
Elsevier Science
Year
1994
Tongue
English
Weight
626 KB
Volume
16
Category
Article
ISSN
1350-4533

No coin nor oath required. For personal study only.

✦ Synopsis


Rate and rate variation algorithms used by implantable devices

designed for management of life-threatening arrhythmias have majo-r limitations in separating physiologic sinus tachycardia (ST) from pathologic ventricular tachycardia (VT) requiring therapy. These algorithms presently utilize criteria such as simple heart rate, stability of rate, or derivative of rate (sudden onset) which assumes a gradual onset for ST and an abrupt onset for VT. An alternative method employing median filtering was designed, tested, and compared to a preoiously published sudden onset rate algorithm using the same data set for analysis of performance. In 50 patients, the onset of ST during exercise and onset of VT were analysed. To accommodate occasional outlying intervals which might affect rate derived by aueraging, a five-cycle median filter was used to smooth heart rate. Results from using a Fxed-interval' or a 'percent' change in the median gave better discrimination of ST and VT than previously published tfixed-interval' or 'percent' change algorithms. The superiority of median filtering perf ormance was validated b statistical measures.