𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Development of a Resonator with Automatic Tuning and Coupling Capability to Minimize Sample Motion Noise for in Vivo EPR Spectroscopy

✍ Scribed by Guanglong He; Sergey Petryakov; Alexandre Samouilov; Michael Chzhan; Periannan Kuppusamy; Jay L. Zweier


Publisher
Elsevier Science
Year
2001
Tongue
English
Weight
327 KB
Volume
149
Category
Article
ISSN
1090-7807

No coin nor oath required. For personal study only.

✦ Synopsis


EPR spectroscopy has been applied to measure free radicals in vivo; however, respiratory, cardiac, and other movements of living animals are a major source of noise and spectral distortion. Sample motions result in changes in resonator frequency, Q, and coupling. These instabilities limit the applications that can be performed and the quality of data that can be obtained. Therefore, it is of great importance to develop resonators with automatic tuning and automatic coupling capability. We report the development of automatic tuning and automatic coupling provisions for a 750-MHz transversely oriented electric field reentrant resonator using two electronically tunable high Q hyperabrupt varactor diodes and feedback loops. In both moving phantoms and living mice, these automatic coupling control and automatic tuning control provisions resulted in an 8- to 10-fold increase in signal-to-noise ratio.