Determination of sterilization effectiveness by measuring bacterial growth in a biological indicator through firefly luciferase determination of ATP
✍ Scribed by Webster, Joann J. ;Walker, Billy G. ;Ford, Sharon R. ;Leach, Franklin R.
- Publisher
- John Wiley and Sons
- Year
- 1988
- Weight
- 372 KB
- Volume
- 2
- Category
- Article
- ISSN
- 0884-3996
No coin nor oath required. For personal study only.
✦ Synopsis
A bioluminescence procedure for measurement of microbial ATP allows a rapid determination of the effectiveness of autoclave sterilization. This determination is achieved faster than detection of acid production in a biological indicator via a pH indicator. Bacterial outgrowth from spores on test strips of the biological indicator was detected by measurement of ATP using the firefly luciferase reaction. A measureable increase in ATP was found after 5 hours of incubation of a biological indicator that had been treated under sterilizing conditions that produced 75% sterility of the biological indicator as measured by acid production. This is a marked improvement over the 24-48 hours of incubation currently required.