𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Determination of oxalate in beer by zone electrophoresis on a chip with conductivity detection

✍ Scribed by Marián Masár; Mária Žúborová; Dušan Kaniansky; Bernd Stanislawski


Publisher
John Wiley and Sons
Year
2003
Tongue
English
Weight
592 KB
Volume
26
Category
Article
ISSN
1615-9306

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

The use of a poly(methylmethacrylate) capillary electrophoresis chip, provided with a high sample load capacity separation system (a 8500 nL separation channel combined with a 500 nL sample injection channel) and a pair of on‐chip conductivity detectors, for zone electrophoresis (ZE) determination of oxalate in beer was studied. Hydrodynamic and electroosmotic flows of the solution in the separation compartment of the chip were suppressed and electrophoresis was a dominant transport process in the separations performed on the chip. A low pH of the carrier electrolyte (3.8), implemented by aspartic acid and bis‐tris propane, provided an adequate selectivity in the separation of oxalate from anionic beer constituents and, at the same time, also a sufficient sensitivity in its conductivity detection. Under our working conditions, this anion could be detected at a 0.5 μmol/L concentration also in samples containing chloride (a major anionic constituent of beer) at a 1800 higher concentration. Such a favorable analyte/matrix concentration ratio made possible accurate and reproducible [typically, 2–5% relative standard deviation (RSD) values of the peak areas of the analyte in dependence on its concentration in the sample] determination of oxalate in 500 nL volumes of 20–50‐fold diluted beer samples. Short analysis times (about 200 s), minimum sample preparation, and reproducible migration times of this analyte (0.5–1.0% RSD values) were characteristic for ZE on the chip.


📜 SIMILAR VOLUMES


Zone electrophoresis separation of perfl
✍ Marián Masár; Lena Wójcik; Dušan Kaniansky; Marek Trojanowicz 📂 Article 📅 2005 🏛 John Wiley and Sons 🌐 English ⚖ 450 KB 👁 1 views

## Abstract Perfluorinated carboxylic acids (PFCAs), amphiphiles of anthropogenic origin, are spread worldwide throughout the environment. This work deals with their zone electrophoresis (ZE) separation on a chip with coupled columns and integrated conductivity detection. Analogies with the electro

Determination of ammonium in wastewaters
✍ Milan Luc; Pavol Kruk; Marián Masár 📂 Article 📅 2011 🏛 John Wiley and Sons 🌐 English ⚖ 175 KB 👁 1 views

## Abstract Analytical potentialities of a chip‐based CE in determination of ammonium in wastewaters were investigated. CZE with the electric field and/or ITP sample stacking was performed on a column‐coupling (CC) chip with integrated conductivity detectors. Acetate background electrolytes (pH ∼3)