✦ Synopsis


A highly sensitive and specific method is described for the determination of methylcyclopentadienylmanganese tricarbonyl (MMT) in gasoline and environmental samples by gas chromatography-plasma atomic emission detection (GC-AED). The procedure is simple, rapid and free from interferences. The absolute limit of detection is 0.5 fg of MMT (0.5ϫ 10 Ϫ 15 g) expressed as Mn. The precision for replicate injections (n = 5) of an MMT solution (10 pg Mn) was 2.8% relative standard deviation (RSD). Applications of the method to analysis of gasoline (petrol), and preliminary results of the analyses of environmental samples such as air and roadside dirt, are given.