๐”– Bobbio Scriptorium
โœฆ   LIBER   โœฆ

Determination of l-glutamate by amperometric flow-injection analysis using immobilized glutamate oxidase: manifold for simultaneous detection of component signal and blank signal

โœ Scribed by Kiyoshi Matsumoto; Koji Sakoda; Yutaka Osajima


Publisher
Elsevier Science
Year
1992
Tongue
English
Weight
388 KB
Volume
261
Category
Article
ISSN
0003-2670

No coin nor oath required. For personal study only.

โœฆ Synopsis


L-Glutamate oxldase was unmoblhzed on Ammo-Cellulofine and used as an enzyme reactor III a flow-mjectlon system The hydrogen peroxrde produced was momtored amperometncally A new configurabon IS described for the detenmnatlon of L-glutamate m food samples for which the matnx provides varymg blank values The peak current was linearly related to L-glutamate concentration m the range 0 01-O 3 mM, and the relatnre standard devlatlon was 1 7% m ten successwe assays at the 0 3 mM level The proposed method was used for seasomng analysis and was compared with an L-glutamate lut (enzymahc, spectrophotometrs) method


๐Ÿ“œ SIMILAR VOLUMES