Determination of iron(III) and titanium(IV) as their Solochrome Violet RS complexes by constant-current stripping potentiometry: Part 1. Automated single-point calibration method for iron(III)

โœ Scribed by Daniel Jagner; Lars Renman; Svana H. Stefansdottir


Publisher
Elsevier Science
Year
1993
Tongue
English
Weight
899 KB
Volume
281
Category
Article
ISSN
0003-2670

No coin nor oath required. For personal study only.

โœฆ Synopsis


An automated constant-current stripping potentiometric method for the determination of iron(III) as its Solochrome Violet RS complex is described. The instrumentation is based on a stripping potentiometry unit, a sample changer, a burette system and a conventional three-electrode system with a mercury film electrode. The method can be utilised over a large concentration range [O-500 pg 1-l iron(III)], by an automated adaption of the experimental conditions to the concentrations found in the samples. A single-point calibration procedure with calibration for three different concentration ranges is used. The possibility for normaliiation of the electrode response to compensate for variations in hydrodynamic conditions and electrode area is demonstrated. Using this normalisation, the reproducibility is better than 5% for samples at the 15 pg 1-l iron level. For higher concentrations (65 and 300 Fg l-l), the corresponding values are 2.9 and 3.5%, respectively. The applicability of the method for the determination of titanium(N) is discussed. The method is applied to the determination of iron(III) in tap water and rain water samples.


๐Ÿ“œ SIMILAR VOLUMES


Determination of iron(III) and titanium(
โœ Daniel Jagner; Lars Renman; Svana H. Stefansdottir ๐Ÿ“‚ Article ๐Ÿ“… 1993 ๐Ÿ› Elsevier Science ๐ŸŒ English โš– 655 KB

## Ahstraet An automated method for the determination of iron and titanium(IV) as their Solochrome Violet RS complexes by constant-current stripping potentiometry and partial least-squares (PLS) regression calibration is described. 'Ihe multivariate calibration procedure overcomes the problem of c