𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Determination of five protein analytes using the beckman array™ and the behring nephelometer system

✍ Scribed by Susan B. Schotters; James H. McBride; Denis O. Rodgerson; Sharon Higgins; Marilyn Pisa


Publisher
John Wiley and Sons
Year
1988
Tongue
English
Weight
359 KB
Volume
2
Category
Article
ISSN
0887-8013

No coin nor oath required. For personal study only.

✦ Synopsis


The Beckman Array and the Behring Ne-meter offered a choice of single or multiphelometer offer the clinical laboratory point calibration, retrieval of raw data, ease completely automated discrete protein of operation, more flexibility in test paramanalysis. To select one of these systems, eters, and choice of calibrator. Interferwe compared five protein analytes: IgG, ence from lipemia frequently caused IgA, IgM, transferrin, and prealbumin. In variable results requiring pretreatment of addition to automation, we wanted to the sample. Both instruments demonachieve better precision than our present strated excellent precision, linearity, and method and have available a wider range correlated adequately with the Hyland of test selection. The Beckman Array gave Laser Nephelometer PDQ. The Array dem-2-week stability of calibration, analyzed li-onstrated slightly better slope values than pemic samples without significant prob-those of the Behring Nephelometer with lems, and had direct manufacturer's the exception of prealbumin analysis. service available. The Behring Nephelo


📜 SIMILAR VOLUMES