Determination of chromium (VI) by flow-injection spectrophotometry coupled with on-line microcolumn preconcentration

✍ Scribed by Wanhong Ma; Ruxiu Cai; Danhua Chen


Publisher
John Wiley and Sons
Year
1999
Tongue
English
Weight
130 KB
Volume
11
Category
Article
ISSN
0895-7533

No coin nor oath required. For personal study only.

✦ Synopsis


A rapid and sensitive method is described for flow-injection spectrophotometric determination of trace chromium (VI) in natural water. Preconcentration and elution of the trace analyte are achieved through quantitative and rapid adsorption on the surface of nanometer-size TiO 2 (anatase) powders and followed by the quantitative and reproducible elution of Cr(VI) using 2.0 M HCl as an eluent. The eluted Cr(VI) reacts with 1,5-diphenylcarbazide (DPCB) in the reaction coil to form a color complex, which is then detected by a spectrophotometer at the wavelength of 540 nm. The preconcentration factor is 55 for the sample volume of 25 mL, and the detection limit is 0.8 ng/mL Cr(VI). Under the optimized conditions, the analysis of Cr(VI) can be achieved at the rate of 6 samples/h for 0.001 lg/mL Cr(VI) or 20 samples/h for 0.1 lg/mL Cr(VI). The relative standard deviations are in the range of 0.37-3.35% (n ‫ס‬ 5).


📜 SIMILAR VOLUMES


Copper determination by flow injection o
✍ Akbar Ali; Hong Shen; Xue Gu; Guangming Xu; Yingxue Ye; Xuefeng Yin 📂 Article 📅 2000 🏛 John Wiley and Sons 🌐 English ⚖ 144 KB

1,10 Phenanthroline (phen) has been tested as a complexing agent, which is more stable and selective than sodium-diethyldithiocarbamate (DDTC) and ammonium-pyrrolidinedithiocarbamate (APDC), for on-line preconcentration of copper on RP-C 18 material in a microcolumn with a flow injection-flame atomi

Robotic sample pretreatment coupled with
✍ Alfredo Velasco-Arjona; María Dolores Luque de Castro; Elissaveta Ivanova; Fredd 📂 Article 📅 1998 🏛 John Wiley and Sons 🌐 English ⚖ 172 KB 👁 2 views

An automated robotic method for sample pretreatment was developed for copper determination in environmental and biological materials at the low microgram per gram level. The robotic station performing the preliminary operations as weighing of sample and acid digestion was coupled with the flowinject