Detection of hard X-rays using superconducting tunnel junctions
β Scribed by Sato, H; Takizawa, Y; Shiki, S; Otani, C; Kurakado, M; Shimizu, H.M
- Publisher
- Elsevier Science
- Year
- 2004
- Tongue
- English
- Weight
- 263 KB
- Volume
- 520
- Category
- Article
- ISSN
- 0168-9002
No coin nor oath required. For personal study only.
β¦ Synopsis
Development of cryogenic detectors for hard X-rays and a-particles using superconducting tunnel junctions has been started. Detector chips with series STJ arrays have been fabricated and experiments for evaluation of the chip properties have been carried out. 0.77% of energy resolution in full-width at half-maximum for 5:48 MeV aparticles has been obtained. To the contrary, energy resolutions for hard X-rays have been poor because of bad signalto-noise ratio.
π SIMILAR VOLUMES
An energy resolution of 1E = 12 eV (FWHM) was measured for the 55 Mn Ka 1 line (E = 5.90 keV) using a single superconducting Al/Al x O y /Al tunnel junction. The total detector area of 100Γ100 Β΅m was illuminated, while a slit mask of about 150 Β΅m width was used to shadow the detector leads and subst