Detection limit of ion-selective bulk optodes and corresponding electrodes
✍ Scribed by Eric Bakker; Michael Willer; Ernö Pretsch
- Book ID
- 102982854
- Publisher
- Elsevier Science
- Year
- 1993
- Tongue
- English
- Weight
- 635 KB
- Volume
- 282
- Category
- Article
- ISSN
- 0003-2670
No coin nor oath required. For personal study only.
✦ Synopsis
The detection limit of a highly selective Pb2+ -sensitive optode based on a solvent polymeric membrane phase is compared to the corresponding Pb*+ -electrode. For this purpose, a practical definition of the detection limit for optodes is presented. Whereas the optode phase has an incorporated Pb*+-selective ionophore and H+-selective chromoionophore, the ion-selective electrode has no chromoionophore, but otherwise equal composition. It is shown that both optode and electrode are responding to the free ion activity of Pb'+ in the sample, which may be buffered by the use of suitable ligands such as NTA and EDTA. The detection limit of the optode (10-13.1 M) has been found to be much lower than that of the electrode (10-9.2 MI, but with sample activities of less than 1O-11.5 M, response times of more than 60 h have been observed with EDTA as metal ion buffer. This is possibly due to a slow decomplexation step of the EDTA-Pb'+ complex.
📜 SIMILAR VOLUMES