𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Design and performance of an in-line surface-induced dissociation device in a four-sector mass spectrometer

✍ Scribed by Kevin L. Schey; David A. Durkin; Kelly R. Thornburg


Publisher
Elsevier Science
Year
1995
Tongue
English
Weight
740 KB
Volume
6
Category
Article
ISSN
1044-0305

No coin nor oath required. For personal study only.

✦ Synopsis


A new in-line surface-induced dissociation device has been designed and characterized in a high performance four-sector tandem mass spectrometer. The design incorporates a target electrode parallel to the ion beam axis and an angled deflector plate (45Β° relative to the ion beam) to provide large collision angles. In addition, an extraction electrode (parallel to the target electrode) is employed to efficiently extract product ions from the target surface. Results obtained with this device indicate high internal energy deposition (up to 16. 3 eV) as measured with the thermometer ions W(CO) 6 (+Β·) and Si(C2H5)4/+Β·, as evidenced by extensive dissociation of the refractory pyrene molecular ion, and as indicated by the b 3/y 2 ratio in the product ion spectrum of leucine enkephalin. High resolution provided by the four-sector instrument for both precursor ions and product ions allows the observation of previously unobserved dissociation products in the surface-induced dissociation spectra of Si(C2H5)4/+Β· and novel ion-surface reaction products in spectra of W(CO)6/+Β· ions after collisions with hydrocarbon-covered surfaces. Both hydrogen atom and hydrocarbon abstraction products are observed. The dissociation efficiencies measured with the in-line device are approximately 1% when hydrocarbon-coated surfaces are used and increase fivefold with a fluorinated surface.


πŸ“œ SIMILAR VOLUMES


Surface-induced dissociation in tandem q
✍ Vicki H. Wysocki; Jian-Mei Ding; Jennifer L. Jones; John H. Callahan; Fred L. Ki πŸ“‚ Article πŸ“… 1992 πŸ› Elsevier Science 🌐 English βš– 558 KB

Three different devices that-can be used for surfac&d~ced d&o&ion (SID) m tandem quadrupole instruments are compared here. The designs were compared by examining the fragmentation of several compounds including benzene, W(CO),, and (CH3)4N+. These studies show that SID can be readily implemented on

Experimental aspects of the collision-in
✍ Jozsef Bordas-Nagy; Dominique Despeyroux; Keith R. Jennings; Simon J. Gaskell πŸ“‚ Article πŸ“… 1992 πŸ› John Wiley and Sons 🌐 English βš– 546 KB

The collision-induced decomposition (CID) of ions in the collision cell of a four-sector tandem mass spectrometer (EBEB) was studied as a function of cell potential and collision gas over the mass range 246-1847. When the cell is at a potential of 2 kV or greater above earth, the loss of light ions

Electrospray ionization tandem mass spec
✍ Jehuda Yinon; Joseph E. McClellan; Richard A. Yost πŸ“‚ Article πŸ“… 1997 πŸ› John Wiley and Sons 🌐 English βš– 225 KB πŸ‘ 2 views

## Electrospray ionization (ESI) tandem mass spectrometry collision-induced dissociation (CID) fragmentation processes of a series of nitroaromatic, nitramine and nitrate ester explosives were studied in the negative-ion mode using daughter-ion, parent-ion and neutral loss scans. 2,4,6-Trinitrotolu

An automated preparation device for the
✍ Soledad CΓ‘rdenas; Mercedes Gallego; Miguel ValcΓ‘rcel πŸ“‚ Article πŸ“… 1997 πŸ› John Wiley and Sons 🌐 English βš– 217 KB πŸ‘ 2 views

An on-line, solid-phase extraction gas chromatography/mass spectrometry method for the automated pretreatment of biological fluids is proposed. Several drugs of abuse were used as model compounds. The analytes are trapped on a home-made RP-C 18 column and subsequently eluted with chloroform. The fra