A circuit is described for a three-electrode controlled alternating current oscilloscopic polarograph . The essential features of this instrument are (1) use of an auxiliary reference electrode to facilitate calibration of the potential axis, (2) automatic it compensation, (3) no restriction on curr
Description of a three-dimensional polarograph
β Scribed by F. David; H. Ouguenoune; A. Bolyos; N. Papadopoulos
- Publisher
- Elsevier Science
- Year
- 1994
- Tongue
- English
- Weight
- 746 KB
- Volume
- 292
- Category
- Article
- ISSN
- 0003-2670
No coin nor oath required. For personal study only.
β¦ Synopsis
All the mechanistic and kinetic information for an electrochemical reaction lies on the i-E-t
surface. The collection of all these data is almost impossible without a computerized polarograph. This paper describes the construction of such an instrument. For the creation of the software that controls a three-dimensional polarograph, Labview, a special programming language for data flow programming, was used. In this language the programmer designs the block diagram of 'the control program and this block diagram is the actual program. In this way the programming work is considerably simplified. In order to test the instrument, the electrochemical oxidation of the iodide ion on mercury was studied.