๐”– Bobbio Scriptorium
โœฆ   LIBER   โœฆ

Creep rupture behavior of polypropylene suture material and its applications as a time-release mechanism

โœ Scribed by Kusy, R. P. ;Whitely, J. Q.


Publisher
John Wiley and Sons
Year
1983
Tongue
English
Weight
728 KB
Volume
17
Category
Article
ISSN
0021-9304

No coin nor oath required. For personal study only.

โœฆ Synopsis


The controlled failure of polypropylene (PP) sutures is studied via creep rupture tests.

From plots of log time ( I S ) vs. stress (a), linear relationships are generated over the failure times of 1-1000 h. Results show that as a function of stress, the time dependence varies with irradiation dose (15,20,25, and 50 Mrad), irradiation atmosphere (air and vacuum), suture diameter (7-0,6-0,5-0, and 4-0), and test temperature (26 and 37ยฐC). For a given stress, the time to failure is least for the greatest dose in the presence of air and at the highest temperature. When suture loops are wrapped around a small wire sheave, however, failure occurs in the largest suture as much as two decades sooner than the smallest suture studied. Within the limitations stated herein, they are independent of test method, loop diameter, aging, and humidity. Consequently, after irradiation in vacuum and postirradiation heat treatment, the processed material may be stored at room temperature for at least 1 month. Such materials are advocated when the time release of a dental or medical device is required, for example, in the self-activating cleft palate appliance.