Correction for displacement artifacts in 3D phase contrast imaging
✍ Scribed by Per Thunberg; Lars Wigström; Tino Ebbers; Matts Karlsson
- Publisher
- John Wiley and Sons
- Year
- 2002
- Tongue
- English
- Weight
- 351 KB
- Volume
- 16
- Category
- Article
- ISSN
- 1053-1807
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Purpose
To correct for displacement artifacts in 3D phase contrast imaging.
Materials and Methods
A 3D phase contrast pulse sequence was modified so that displacements of velocity measurements were restricted to one direction. By applying a postprocessing method, displaced measurements could be traced back to their accurate positions. Flow studies were performed using a phantom that generated flow through a stenosis, directed oblique relative to the phase and frequency encoding directions. Velocity profiles and streamline visualization were used to compare displaced and corrected velocity data to a reference.
Results
Velocity profiles obtained from the original measurement showed skewed profiles due to the displacement artifact, both at close proximity to the orifice as well as further downstream. After correction, concordance with the reference improved considerably.
Conclusion
The displacement artifact, which restricts the accuracy of phase contrast measurements, can be corrected for using the proposed method. Correction of the phase contrast velocity data may improve the accuracy of subsequent flow analysis and visualization. J. Magn. Reson. Imaging 2002;16:591–597. © 2002 Wiley‐Liss, Inc.
📜 SIMILAR VOLUMES
## Abstract A theoretical framework for understanding acceleration‐induced errors in phase‐contrast magnetic resonance velocity measurements has been developed. An important result of this framework is the interpretation of acceleration‐induced velocity errors as displacement artifacts due to the d
## Abstract ## Purpose To show that accuracy of jet flow representation by magnetic resonance (MR) phase‐contrast (PC) velocity‐encoded (VE) cine imaging is dominated by error terms resulting from the temporal distribution of data, and to present a generally applicable data interpolation‐based app
## Abstract One of the difficulties encountered in ^19^F NMR imaging of fluorinated blood substitutes is that these compounds often exhibit complex multipeak spectra. These peaks result in chemical‐shift artifacts along the readout direction and blurred images. In addition, each peak excites a diff
## Abstract ## Purpose: To evaluate the use of computational fluid dynamic (CFD)‐based magnetic resonance imaging (MRI) simulations to predict the image appearance and velocity measurement of fluid flow in human carotid bifurcation geometries, and to compare the results with images from experiment