𝔖 Bobbio Scriptorium
✦   LIBER   ✦

Correction factors to account for minor sample height variations in gamma-ray spectrometry

✍ Scribed by P. Jodłowski


Publisher
Elsevier Science
Year
2007
Tongue
English
Weight
140 KB
Volume
580
Category
Article
ISSN
0168-9002

No coin nor oath required. For personal study only.

✦ Synopsis


It often happens in gamma-ray spectrometry measurements that there is less sample material than required for the given measuring geometry. One approach to this problem is to stick to the original geometry, though a correction factor is to be introduced to account for the difference in the height of the sample material. This correction factor C h is expressed as the ratio of spectrometer efficiency for the nominal sample height e(h 0 ) to that obtained for the actual height e(h).

The author determined the correction factor C h for several radiation energies E, 81.0, 356.0, 661.7 and 1173.2 keV. Two measurement geometries were considered: a Marinelli beaker 710 cm 3 in volume, and a cylindrical sample 31.5 mm in height. The correction factors were obtained experimentally and by Monte Carlo simulation method for h falling within the range h 0 78 mm. C h values obtained by these two methods are consistent. For EX356 keV, C h value almost does not depend on energy. C h value varies linearly with dh. For Marinelli beaker for EX356 keV, the correction is 0.9% (1.0% for E ¼ 81 keV) for each millimeter of sample height change; for a cylindrical geometry the correction is 1.5% (1.7%). Monte Carlo method was further used to compute C h values for several other cylindrical geometries. The lower the nominal height of the cylindrical samples, the more sensitive the method to sample height variations. The knowledge of the value C h enables us to estimate the uncertainty of the measurements, associated with the sample height uncertainty.