## Abstract __T__~1~‐shortening contrast agents have been used to improve the depiction of coronary arteries with breath‐hold magnetic resonance angiography (MRA). The spatial resolution and coverage are limited by the duration of the arterial phase of the contrast media passage. In this study we i
Contrast-enhanced whole-heart coronary MRI with bolus infusion of gadobenate dimeglumine at 1.5 T
✍ Scribed by Peng Hu; Jonathan Chan; Long H. Ngo; Jouke Smink; Beth Goddu; Kraig V. Kissinger; Lois Goepfert; Thomas H. Hauser; Neil M. Rofsky; Warren J. Manning; Reza Nezafat
- Publisher
- John Wiley and Sons
- Year
- 2010
- Tongue
- English
- Weight
- 968 KB
- Volume
- 65
- Category
- Article
- ISSN
- 0740-3194
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
We sought to investigate the T~1~ kinetics of blood and myocardium after three infusion schemes of gadobenate dimeglumine (Gd‐BOPTA) and subsequently compared contrast‐enhanced whole‐heart coronary MRI after a bolus Gd‐BOPTA infusion with nonenhanced coronary MRI at 1.5 T. Blood and myocardium T~1~ was measured in seven healthy adults, after each underwent three Gd‐BOPTA infusion schemes (bolus: 0.2 mmol/kg at 2 mL/sec, hybrid: 0.1 mmol/kg at 2 mL/sec followed by 0.1 mmol/kg at 0.1 mL/sec, and slow: 0.2 mmol/kg at 0.3 mL/sec). Fourteen additional subjects underwent contrast‐enhanced coronary MRI with an inversion‐recovery steady‐state free precession sequence after bolus Gd‐BOPTA infusion. Images were compared with nonenhanced T~2~‐prepared steady‐state free precision whole‐heart coronary MRI in signal‐to‐noise ratio, contrast‐to‐noise ratio, depicted vessel length, vessel sharpness, and subjective image quality. Bolus and slow infusion schemes resulted in similar T~1~ during coronary MRI, whereas the hybrid infusion method yielded higher T~1~ values. A bolus infusion of Gd‐BOPTA significantly improved signal‐to‐noise ratio, contrast‐to‐noise ratio, depicted coronary artery length, and subjective image quality, when all segments were collectively compared but not when compared segment by segment. In conclusion, whole‐heart steady‐state free precision coronary MRI at 1.5 T can benefit from a bolus infusion of 0.2 mmol/kg Gd‐BOPTA. Magn Reson Med, 2011. © 2010 Wiley‐Liss, Inc.
📜 SIMILAR VOLUMES