Construction of a thermodesorption injector for gas chromatography
✍ Scribed by Wolfgang Schröder
- Publisher
- John Wiley and Sons
- Year
- 2011
- Tongue
- English
- Weight
- 302 KB
- Volume
- 34
- Category
- Article
- ISSN
- 1615-9306
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
A detailed description for the do‐it‐yourself construction of a high‐performance thermodesorption GC injector (HPTI) is presented. The injector consists of a heated desorption chamber, a cooled loop for focusing the desorbed volatiles as well as a heating control for sample injection into the separation column. The gas flow is controlled by valve switching during the steps of the operation. Two versions of the injector function either for the injection of highly volatile compounds or for compounds with extremely low volatility, e.g. lubricants. The sophisticated gas flow management allows injection of highly concentrated samples without memory effects in a sample gas split mode, while traces may be injected with virtually 1:1 sample transfer. A wide range of adsorbents appropriate for the compounds can be chosen due to homemade adsorbent tubes and the achievability of high desorption temperature. Solutions may be injected from a carrier material after solvent evaporation. The HPTI fulfills all multipurpose GC injector requirements and with the support of downloadable instruction files can be built by anyone working with conventional GC devices.
📜 SIMILAR VOLUMES
Gas chromatographyfion-trap mass spectrometry in the electron ionization (EI) mode, with a cooled temperature-programmable injector, has been used for the analysis of a series of thermolabile carbamate and phenylurea pesticides. The mass spectra were found to be in good agreement with 70eV solid-pro
This article describes a method to determine standard pesticides in a model solution based on very large volume injection in capillary gas chromatogra-Ž . phy using a temperature-programmable injector PTV . The influences of sample volume, speed of sample introduction, flow rate during sampling, and
## Abstract On‐column injection into a hot gas chromatograph oven involves the danger that the bottom part of the needle is warmed up above the boiling point of the solvent in the sample. Consequently parts of the sample are evaporated out of the needle instead of being introduced as a liquid into