Computational comparison of reamed versus unreamed intramedullary tibial nails
✍ Scribed by María José Gómez-Benito; Pere Fornells; José Manuel García-Aznar; Belén Seral; Fernando Seral-Iñnigo; Manuel Doblaré
- Book ID
- 102395560
- Publisher
- Elsevier Science
- Year
- 2007
- Tongue
- English
- Weight
- 401 KB
- Volume
- 25
- Category
- Article
- ISSN
- 0736-0266
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
We compared, via a computational model, the biomechanical performance of reamed versus unreamed intramedullary tibial nails to treat fractures in three different locations: proximal, mid‐diaphyseal, and distal. Two finite element models were analyzed for the two nail types and the three kinds of fractures. Several biomechanical variables were determined: interfragmentary strains in the fracture site, von Mises stresses in nails and bolts, and strain distributions in the tibia and fibula. Although good mechanical stabilization was achieved in all the simulated fractures, the best results were obtained in the proximal fracture for the unreamed nail and in the mid‐diaphyseal and distal fractures for the reamed nail. The interlocking bolts, in general, were subjected to higher stresses in the unreamed tibial nail than in the reamed one; thus the former stabilization technique is more likely to fail due to fatigue. © 2006 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 25:191–200, 2007