In 22 patients with a diverse range of thoracic abnormalities, T2-weighted magnetic resonance (MR) images of the chest were obtained with electrocardiograph (ECG)-triggered turbo spin-echo (TSE), ECG-triggered conventional spin-echo (CSE), and nontriggered TSE sequences, and the images were compared
Comparison of ultrafast half-Fourier single-shot turbo spin-echo sequence with turbo spin-echo sequences for T2-weighted imaging of the female pelvis
✍ Scribed by Yasuyuki Yamashita; Yi Tang; Yasuko Abe; Katsuhiko Mitsuzaki; Matsumasa Takahashi
- Publisher
- John Wiley and Sons
- Year
- 1998
- Tongue
- English
- Weight
- 778 KB
- Volume
- 8
- Category
- Article
- ISSN
- 1053-1807
No coin nor oath required. For personal study only.
✦ Synopsis
So that we might evaluate the ultrafast half-Fourier single-shot turbo spin-echo (HASTE) sequence in T2-weighted MRI of the female pelvis and compare it with the turbo spin-echo (TSE) sequence, we prospectively studied 60 consecutive females with suspected abnormalities of the pelvis. For all MR examinations, we used a 1.5-T superconductive magnet with a phased array coil. The HASTE sequence was applied with TR/effective TE/echo train = infinity/90/64 and a 128 x 256 matrix (acquisition time: .3 sec/slice), conventional TSE imaging with 3,400 to 5,000/132/15 and a 128 x 256 matrix (mean acquisition time: 2 min 4 sec), and high-resolution TSE imaging with 3,400 to 5,000/132/15 and a 300 x 512 matrix (6 min 4 sec). Although the lesion conspicuity for the HASTE sequence was less than that for the high-resolution TSE sequences, artifacts (including ghosting, bowel motion, susceptibility difference, and chemical shift) were negligible on HASTE images of all patients. The lesion conspicuity for the HASTE sequence was significantly better than for the conventional TSE sequence. In spite of the very short acquisition time, the subjective scoring of the overall image quality for the HASTE sequence was significantly higher than for the conventional TSE sequence (P < .01) and were slightly lower than for the high-resolution TSE sequence. Compared with high-resolution TSE, HASTE provided clearer visualization of large leiomyomas and ovarian tumors but slightly poorer visualization of uterine cancer. In occlusion, HASTE sequence generates higher contrast and is free from motion and chemical shift artifact with much higher time efficacy. Because of limited image resolution, the HASTE sequence should be used when the high-resolution TSE imaging is suboptimal.
📜 SIMILAR VOLUMES
## Abstract The purpose of this study was to quantitatively compare the hepatic contrast characteristics of conventional spin‐echo (CSE) and fast spin‐echo (FSE) sequences with breath‐hold T2‐weighted images acquired with half‐Fourier turbo spin echo (HASTE). Forty‐five patients were examined with
## Abstract Magnetic resonance (MR) imaging with multi‐planar, gated Half‐Fourier Acquisition Single‐Shot Turbo Spin‐Echo (HASTE) imaging was performed to evaluate a woman with dysrhythmia. MR examination for right ventricular dysplasia revealed a Morgagni hernia. The HASTE images depicted well the
## Abstract ## Purpose To optimize and assess the feasibility of a single‐shot black‐blood T2‐weighted spin‐echo echo‐planar imaging (SSBB‐EPI) sequence for MRI of the liver using sensitivity encoding (SENSE), and compare the results with those obtained with a T2‐weighted turbo spin‐echo (TSE) seq
## Abstract Single shot (SS) rapid acquisition with relaxation enhancement (RARE) and half Fourier SS‐RARE (HFSS‐RARE, HASTE, or SS‐FSE) sequences allow ultrafast imaging acquisition and generate high imaging quality. Images can be acquired within a very short time, without artifacts from physiolog
The purpose of this study was to compare the relative usefulness of multishot turbo spin echo (TSE) and half-Fourier single-shot turbo spin echo (HASTE) for determination of optimal breath-hold fast T2-weighted technique in terms of lesion detection, lesion-to-liver contrast-to-noise ratio (CNR), an