Comparison of two automated nephelometers
โ Scribed by Kenneth Emancipator; Ronald J. Elin; Thomas A. Fleisher
- Book ID
- 102307777
- Publisher
- John Wiley and Sons
- Year
- 1992
- Tongue
- English
- Weight
- 451 KB
- Volume
- 6
- Category
- Article
- ISSN
- 0887-8013
No coin nor oath required. For personal study only.
โฆ Synopsis
Abstract
The intercalibration precision and linearity were determined for two representative analytes, apolipoprotein A1 (apo A1) and immunoglobulin G (IgG), on the Beckman Arrayยฎ and the Behring Nephelometer 100ยฎ (BNโ100). For two of nine samples analyzed, poor precision was observed for IgG with the Array. The poor precision for these two samples is attributed to large systematic shifts. A statistical nonโlinearity was observed for apo A1 with the BNโ100, but this nonโlinearity does not exclude use of this assay for clinical diagnosis. Method comparisons were done for 12 analytes: apo A1, apo B, IgG, IgA, IgM, IgE, C3, C4, albumin, Cโreactive protein, alphaโ1โantitrypsin, and ceruloplasmin. Thjese comparisons showed proportional biases of >10% for seven of 12 analytes. Furthermore, correlation coefficients were <0.96 for seven of 12 analytes. We conclude that comparing results obtained from th two different nephelometers is of only limited value for the individual patient. ยฉ 1992 WileyโLiss, Inc.