Comparison of the stability of various internal fixators used in the treatment of osteochondritis dissecans—a mechanical model

✍ Scribed by Moreno Morelli; Philippe Poitras; Valentine Grimes; David Backman; Geoffrey Dervin


Publisher
Elsevier Science
Year
2007
Tongue
English
Weight
202 KB
Volume
25
Category
Article
ISSN
0736-0266

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

The purpose of this study was to determine what characteristics of fixation devices used in the treatment of osteochondritis dissecans (OCD) contribute to improved stability to resist shear loading. An OCD model was designed using rigid polyurethane foam. Each specimen consisted of two components, an osteochondral fragment and a corresponding defect. A total of 40 specimens were prepared and assigned to one of four groups: control (no extrinsic stabilizer); two 2‐mm‐diameter Kirschner wires (K‐wires), 40 mm in length; one threaded washer and a 28‐mm screw; and one threaded washer and a 38 mm screw. Each specimen was mounted onto an Iosipescu shear test fixture and subjected to shear loads at a pseudo‐static displacement rate of 0.075 mm/s. All groups demonstrated some stability; controls were significantly less stable than all other groups. The group with the threaded washer and 38‐mm screw demonstrated the greatest stability (p < 0.001), and no difference was noted between the K‐wire and 28‐mm screw groups. These results suggest that, in this OCD model, friction conferred some intrinsic stability to resist loads in shear. However, stability was improved with the use of long implants that compressed the fragments together. © 2007 Orthopaedic Research Society. Published by Wiley Periodicals, Inc. J Orthop Res 25:495–500, 2007


📜 SIMILAR VOLUMES


Model for estimating the threshold mecha
✍ Allison Zemek; Rohit Garg; Brian J. F. Wong 📂 Article 📅 2010 🏛 John Wiley and Sons 🌐 English ⚖ 193 KB

## Abstract ## Objectives/Hypothesis: Characterizing the mechanical properties of structural cartilage grafts used in rhinoplasty is valuable because softer engineered tissues are more time‐ and cost‐efficient to manufacture. The aim of this study is to quantitatively identify the threshold mechan

A comparison of two schedules of cytosin
✍ Slevin, M. L. ;Rohatiner, A. Z. S. ;Dhaliwal, H. S. ;Henry, G. P. ;Bell, R. ;Lis 📂 Article 📅 1982 🏛 John Wiley and Sons 🌐 English ⚖ 368 KB 👁 1 views

The most effective schedule of cytosine arabinoside (Ara-C) administration remains controversial and is further confused by the use of combination chemotherapy. Two remission induction regimens comprising adriamycin, Ara-C, and 6-thioguanine have been compared in patients with acute myelogenous leuk