## Abstract ## Purpose To validate a noninvasive quantitative MRI technique, the K~i~ perfusion method, for myocardial perfusion in humans using ^13^N‐ammonia PET as a reference method. ## Materials and Methods Ten healthy males (64 ± 8 years) were examined with combined PET and MRI perfusion im
Comparison of MRI and positron emission tomography for measuring myocardial perfusion reserve in healthy humans
✍ Scribed by Jussi P. Pärkkä; Pekka Niemi; Antti Saraste; Juha W. Koskenvuo; Markku Komu; Vesa Oikonen; Jyri O. Toikka; Tuomas O. Kiviniemi; Juhani Knuuti; Hajime Sakuma; Jaakko J. Hartiala
- Publisher
- John Wiley and Sons
- Year
- 2006
- Tongue
- English
- Weight
- 217 KB
- Volume
- 55
- Category
- Article
- ISSN
- 0740-3194
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
Myocardial perfusion reserve (MPR, defined as the ratio of the maximum myocardial blood flow (MBF) to the baseline) is an indicator of coronary artery disease and myocardial microvascular abnormalities. First‐pass contrast‐enhanced magnetic resonance imaging (CE‐MRI) using gadolinium (Gd)‐DTPA as a contrast agent (CA) has been used to assess MPR. Tracer kinetic models based on compartmental analysis of the CA uptake have been developed to provide quantitative measures of MBF by MRI. To study the accuracy of Gd‐DTPA first‐pass MRI and kinetic modeling for quantitative analysis of myocardial perfusion and MPR during dipyridamole infusion, we conducted a comparison with positron emission tomography (PET) in 18 healthy males (age = 40 ± 14 years). Five planes were acquired at every second heartbeat with a 1.5T scanner using a saturation recovery turboFLASH sequence. A perfusion‐related parameter, the unidirectional influx constant (K~i~), was computed in three coronary artery territories. There was a significant correlation for both dipyridamole‐induced flow (0.70, P = 0.001) and MPR (0.48, P = 0.04) between MRI and PET. However, we noticed that MRI provided lower MPR values compared to PET (2.5 ± 1.0 vs. 4.3 ± 1.8). We conclude that MRI supplemented with tracer kinetic modeling can be used to quantify myocardial perfusion. Magn Reson Med, 2006. © 2006 Wiley‐Liss, Inc.
📜 SIMILAR VOLUMES