Comparison between the enzymatic vitros assay for creatinine determination and three other methods adapted on the olympus analyzer

✍ Scribed by S. Badiou; A.M. Dupuy; B. Descomps; J.P. Cristolead


Book ID
102311356
Publisher
John Wiley and Sons
Year
2003
Tongue
English
Weight
178 KB
Volume
17
Category
Article
ISSN
0887-8013

No coin nor oath required. For personal study only.

✦ Synopsis


Abstract

To evaluate the relationship between the enzymatic Vitros assay for creatinine determination and other methods, we determined creatinine concentration in 400 heparin samples. Plasma creatinine level was successively determined on the Vitros 750 analyzer (Johnson & Johnson Co., Rochester, NY) and on the Olympus AU2700 analyzer (Olympus France, Rungis, France), using three reagents in the same assay: Olympus‐Jaffé and two enzymatic commercial kits—Crea Plus (Roche Diagnostics, Meylan, France) and Enzymatic Creatinine (Randox, Mauguio, France). Comparison of Jaffé and enzymatic measurements of plasma creatinine levels revealed a high correlation when considering all values ranged from 20–1,000 μmol/l (r > 0.99). However, for values <60 μmol/l, enzymatic reagents provided best results. Enzymatic methodology is a better clinical choice for the accurate measurement of creatinine, especially for neonates, pediatrics, and hematology units. Because analytical performance and the costs of Randox creatinine were satisfactory for our laboratory, this method, adapted on the Olympus analyzer, was chosen for routine determination of creatinine levels. According to the Valtec protocol (8), no interferences such as hemolysis, lipemia, or bilirubin were detected for Enzymatic Creatinine Randox. J. Clin. Lab. Anal. 17:235–240, 2003. © 2003 Wiley‐Liss, Inc.


📜 SIMILAR VOLUMES