Clinical wear of 63 ultrahigh molecular weight polyethylene acetabular components: Effect of starting resin and forming method

✍ Scribed by James, Susan P. ;Lee, Kyla R. ;Beauregard, Guy P. ;Rentfrow, Elisha D. ;McLaughlin, Jeffrey R.


Book ID
101258173
Publisher
John Wiley and Sons
Year
1999
Tongue
English
Weight
342 KB
Volume
48
Category
Article
ISSN
0021-9304

No coin nor oath required. For personal study only.

✦ Synopsis


This study compares the clinical wear rates and implant characteristics of 63 surgically retrieved acetabular components. All components were made by the same manufacturer, implanted by the same surgeon, in articulation against the same type of femoral component, and revised for the same reason; 19 were made from directly compression molded, calcium stearate free, ultrahigh molecular weight polyethylene (UHMWPE) and 44 were made from machined, ram extruded, calcium stearate containing UHMWPE. There were significant differences in wear, duration, and wear rate between the molded (type I) and machined (type II and III) components. Most importantly, the wear rates of type I (molded) components were significantly (p < 0.0001) lower than the wear rates of type II, type III, and type II and III components as a group (all machined). The machined components had wear rates 2.3 times greater than the molded components. The wear rates between the two different groups of machined components (type II and III) were not significantly different. The scanning electron microscope observations did not reveal any major differences in wear mechanisms between the three types of components, although the machined components did show more evidence of brittleness. The molded components were better consolidated (or had less fusion defects) than the ram extruded components.


📜 SIMILAR VOLUMES


Quantifying the effect of resin type and
✍ Willie, Bettina M. ;Ashrafi, Shadi ;Alajbegovic, Sanjin ;Burnett, Trever ;Bloeba 📂 Article 📅 2004 🏛 John Wiley and Sons 🌐 English ⚖ 257 KB

## Abstract Alternative sterilization methods including ethylene oxide, gas plasma, and gamma‐radiation in an inert environment were implemented in the late 1990s, to limit oxidative degradation of ultrahigh molecular weight polyethylene (PE). There was also a simultaneous transition to PE resins t