A voltage impulse generator with a varying repetition time for supplying electron-capture detectors is described. The generator allows the measurement of the detector ionization current with a change of the pulse repetition time (tr) from 1 to 5200 psec in 8.3 min. The generator allows the measureme
Circuit for constant current operation of an electron capture detector
β Scribed by Knighton, W. B.; Grimsrud, E. P.
- Book ID
- 124078132
- Publisher
- American Chemical Society
- Year
- 1982
- Tongue
- English
- Weight
- 242 KB
- Volume
- 54
- Category
- Article
- ISSN
- 0003-2700
No coin nor oath required. For personal study only.
π SIMILAR VOLUMES
THE first edition of the Bibliography of Scientific Literature relating to helium has been revised, and references to the literature published during the years 1919--1921 have been added. As in the first edition, the titles are grouped under numerous headings and so arranged under each heading that
The use of an electron capture detector for the determination of chlorinated insecticide residues in water is described. A linear response of the detector was found for BHC and DDT isomers and metabolites in the range 0.03-l -20 ,ug/ml. Results for the determination of chlorinated insecticide +BHC;
## Abstract Subnanogram detection of steroids has become increasingly important today. One applicable method for gas chromatographic determination of subnanogram quantities of estrogens as halogenated derivatives is electron capture detection. HFBβderivatives of 7 different estrogens were automatic