The determination of molecular mass (MM) data for polydisperse polymers by size exclusion chromatography matrix assisted laser desorption/ionization time-of-flight (SEC/MALDI-TOF) involves the fractionation of samples through an analytical SEC. Selected fractions are then analysed by MALDI-TOF and t
Characterization of polydisperse synthetic polymers by size-exclusion chromatography/matrix-assisted laser desorption/ionization time-of-flight mass spectrometry
β Scribed by Michel W. F. Nielen; Sabine Malucha
- Publisher
- John Wiley and Sons
- Year
- 1997
- Tongue
- English
- Weight
- 228 KB
- Volume
- 11
- Category
- Article
- ISSN
- 0951-4198
No coin nor oath required. For personal study only.
β¦ Synopsis
Synthetic polymers having a polydispersity greater than 1.1 cannot be reliably characterized by matrix-assisted laser desorption/ionization (MALDI) time-of-flight mass spectrometry alone. This complication has been overcome by off-line coupling of size-exclusion chromatography (SEC) with MALDI. Data are presented for a wide variety of synthetic polymers having polydispersities from 1.7 up to 3.0. Polystyrene (15 kDa and 48 kDa), polybutylacrylate (62 kDa), polycarbonate (28 kDa), aromatic polyester resin (8 kDa) and a methyl methacrylate-methacrylic acid copolymer (34 kDa) were injected on a conventional SEC system with refractive index (RI) or UV absorbance detection and 40 fractions collected over the polymer distribution. Ten fractions were analyzed by a MALDI time-of-flight mass spectrometer operated in the continuous extraction linear mode. The m/z thus obtained were used as calibration points for the SEC-RI system and allowed accurate calculation of the molecular weight averages M w , M n , M z and the polydispersity for all the polymers tested. The data for these entirely different polydisperse polymers were in good agreement with reference data from, for example, SEC with intrinsic viscosity detection. In addition, the choices of matrix and cation for individual types of polymers are commented on. The end groups of the polymers were determined from isotopically-resolved oligomer spectra recorded by MALDI time-of-flight in the delayed-extraction reflectron mode. End groups proposed are compared with reference data from, amongst others, NMR and MALDI Fourier transform ion cyclotron resonance-MS measurements.
π SIMILAR VOLUMES
High-resolution Infrared Laser Desorption/Ionization and Matrix-assisted
Matrix-assisted laser desorption/ionization time-of-flight mass spectrometry was used to structurally investigate a synthetic model of lignin, obtained by polymerization of coniferyl alcohol in the presence of horseradish peroxidase and hydrogen peroxide. In contrast to the widespread understanding
A variety of gram-positive and gram-negative intact bacterial cells have been analysed by matrix-assisted laser desorption/ionization mass spectrometry (MALDI-MS) and shown to provide fingerprint mass spectra with discrete peaks being observed over the mass range from 3 to 40 kDa. The spectra show b