The wear of nonirradiated ultra-high molecular weight polyethylene (UHMWPE) was studied with a circularly translating pin-on-disk (CTPOD) device. With this simple device, the wear of the acetabular cup is simulated. Two shapes of a polyethylene pin were used: flat wear faces of 3.0 and 8.9 mm in dia
Characterization and tribology of PEG-like coatings on UHMWPE for total hip replacements
✍ Scribed by Sheryl R. Kane; Paul D. Ashby; Lisa A. Pruitt
- Publisher
- John Wiley and Sons
- Year
- 2009
- Tongue
- English
- Weight
- 913 KB
- Volume
- 9999A
- Category
- Article
- ISSN
- 1549-3296
No coin nor oath required. For personal study only.
✦ Synopsis
Abstract
A crosslinked hydrogel coating similar to poly(ethylene glycol) (PEG) was covalently bonded to the surface of ultrahigh molecular weight polyethylene (UHMWPE) to improve the lubricity and wear resistance of the UHWMPE for use in total joint replacements. The chemistry, hydrophilicity, and protein adsorption resistance of the coatings were determined, and the wear behavior of the PEG‐like coating was examined by two methods: pin‐on‐disk tribometry to evaluate macroscale behavior, and atomic force microscopy (AFM) to simulate asperity wear. As expected, the coating was found to be highly PEG‐like, with ∼83% ether content by x‐ray photoelectron spectroscopy and more hydrophilic and resistant to protein adsorption than uncoated UHMWPE. Pin‐on‐disk testing showed that the PEG‐like coating could survive 3 MPa of contact pressure, comparable to that experienced by total hip replacements. AFM nanoscratching experiments uncovered three damage mechanisms for the coatings: adhesion/microfracture, pure adhesion, and delamination. The latter two mechanisms appear to correlate well with wear patterns induced by pin‐on‐disk testing and evaluated by attenuated total reflection Fourier transform infrared spectroscopy mapping. Understanding the mechanisms by which the PEG‐like coatings wear is critical for improving the behavior of subsequent generations of wear‐resistant hydrogel coatings. © 2009 Wiley Periodicals, Inc. J Biomed Mater Res 2010
📜 SIMILAR VOLUMES
The less intense tissue reaction around metal on metal total hip replacements (THRs) compared to metal on polyethylene (PE) THRs may be explained by the differences in the characteristics of metal wear particles. In this study, transmission electron microscopy was used to study metal wear particles