๐”– Bobbio Scriptorium
โœฆ   LIBER   โœฆ

Catalase activity measured with a micro oxygen electrode in a pressurized reaction vessel

โœ Scribed by Stefan Halbach


Book ID
102984073
Publisher
Elsevier Science
Year
1977
Tongue
English
Weight
617 KB
Volume
80
Category
Article
ISSN
0003-2697

No coin nor oath required. For personal study only.

โœฆ Synopsis


The assembly and the use of a simple airtight pressurized reaction vessel are described for the measurement of catalase activity with a micro oxygen electrode in an optically heterogenous medium. The oxygen concentration is expressed as the ratio of observed current to the current in an air-saturated solution. Thus, an individual standard can be obtained for each measurement and the calibration is less affected by changes in the amplification factor. Different procedures for calibration of the oxygen electrode were compared. Specific activities of crystalline catalase, of red blood cells from humans and from normal or acatalasemic mice, and of liver homogenates from normal or aminotriazole-pretreated rats were determined. The specific catalase activity of human erythrocytes, as found by this method, agrees with that obtained photometrically.